JavaScript seems to be disabled in your browser. For the best experience on our site, be sure to turn on Javascript in your browser.
DesignEN 837-3
Nominal size in mm63, 100, 160
Accuracy class1.6
Scale rangesNS 63: 0 ... 40 mbar to 0 ... 600 mbarNS 100: 0 ... 16 mbar to 0 ... 600 mbarNS 160: 0 ... 2.5 mbar to 0 ... 600 mbaror all other equivalent vacuum or combined pressure and vacuum ranges
Pressure limitationSteady: full scale valueFluctuating: 0.9 x full scale value
Operating temperatureAmbient: -20 ... +60 °CMedium: +100 °C maximum
Temperature effectWhen the temperature of the measuring system deviates from the reference temperature (+20 °C):max. ±0.6 %/10 K of full scale value
Ingress protectionIP 54 per EN 60 529 / IEC 529(with liquid filling IP 65)