• Environmental Compensated Temperature: 14°F to 122°F(-10°C to 50°C ) Operating Temperature*: 14°F to 122°F(-10°C to 50°C ) Storage Temperature: -4°F to 158°F (-20°C to 70°C ) Humidity: <95%
• Pressure Connections 1/4NPT male, 1/4BSP male, M20X1.5 male Other Connections available per request
• Enclosure Case material: SST Wetted parts: 316L SS Dimension: Ø1.3″ x 4.84″ (Ø33mm x 123mm) Weight: 0.99 lb (0.4kg)
• Warranty: 1 Year
• Pressure Range and Accuracy
Gauge Pressure [1]
P/N
Pressure Range
Media [2]
Accuracy(%FS)
Burst Pressure
(psi)
V15
-15
-1.0
G
0.02
3×
GP2
2
0.16
0.05
GP5
5
0.35
GP10
10
0.7
GP15
15
1.0
GP30
30
2.0
GP50
50
3.5
G , L
GP100
100
7.0
GP300
300
20
GP500
500
35
GP600
600
40
GP1K
1,000
70
2×
GP2K
2,000
140
GP3K
3,000
200
GP5K
5,000
350
GP10K
10,000
700
Note: [1]. Sealed gauge pressure for above 1000 psi [2]. G=Gas, L=Liquid
Compound Pressure
Media
(bar)
CP2
±2
±0.16
CP5
±5
±0.35
CP10
±10
±0.7
CP15
±15
±1
CP30
﹣15 to 30
﹣1 to 2
CP100
﹣15 to 100
﹣1 to 7
G, L
CP300
﹣15 to 300
﹣1 to 20
Absolute Pressure
AP5
0.1
AP10
AP15
AP30
AP50
AP100
0.05 (0.1)
AP300
AP500
AP1K
AP3K
AP5K
Differential Pressure
(inH2O)
(mbar)
DP1
±2.5
100×
DP2
±5.0
DP5
50×
DP10
±25
20×
DP20
±20
±50
DP30
±30
±75
15×
DP50
±160
DP150
±150
±350
DP300
±300
±700
* One year accuracy (including yearly stability) except DP1 and DP2 range which is 0.05%FS calibration accuracy and 0.05%FS yearly stability.