Additel ADT 223A Documenting Multifunction Process Calibrator with HART

The Additel ADT 223A documenting multifunction calibrator with HART is a compact, rugged and easy to use hand held device for sourcing, simulating and measuring pressure, temperature and electrical signals. With HART communication capabilities, the Additel ADT 223A is ideal for calibrating, maintaining and troubleshooting HART instrumentation.
The Additel ADT 223A documenting multifunction calibrator with HART is a compact, rugged and easy to use hand held device for sourcing, simulating and measuring pressure, temperature and electrical signals. With HART communication capabilities, the Additel ADT 223A is ideal for calibrating, maintaining and troubleshooting HART instrumentation.
More Information
Features
  • Measure and source 13 T/Cs, 11 RTDs, mV, mA, Ohms, frequency and pulses, with accuracy to 0.01% of reading
  • Pressure via
  • external pressure modules
  • up to 70,000kPa
  • 24V loop power
  • HART communication capability
  • Task management capability, upload, download and document
  • Easy to use smartphone type menu and interface
  • Rugged design for use in harsh environments
Specifications

• Electrical Specifications

Measurement Accuracy
Range
Resolution
Accuracy 
Voltage DC
±75.0000 mV
±30.0000 V
0.11µV
0.1mV
0.01%RD+3.75 μV
0.01%RD+1.5 mV
Current DC
±30.0000 mA
0.1µA
0.01%rdg + 1.5 μA
Resistance
Two-wire
0 to 400.000 Ω
1mΩ
0.02%RD + 0.02 Ω
Three-wire
0 to 400.000 Ω
1mΩ
0.02%RD + 0.02 Ω
Four-wire
0 to 400.000 Ω
1mΩ
0.01%RD + 0.02 Ω
Two-wire
0 to 400.000 Ω
10mΩ
0.02%RD + 0.2 Ω
Three-wire
0 to 400.000 Ω
10mΩ
0.02%RD + 0.2 Ω
Four-wire
0 to 400.000 Ω
10mΩ
0.01%RD + 0.2 Ω
Frequency
1 to 50000.0 Hz
0.1Hz
0.005%RD + 1 Hz
Pulse
0 to 999999
1
N/A
Limit Switch
If the switch has detective voltage, its range is +3V to +24V
Source Accuracy
Voltage DC
-10.000 to 75.000mV 0 to 12.0000 V
1μV
0.1mV
0.02%RD + 4.25 μV
0.02%RD + 0.6 mV
Current DC
0 to 22.000 mA
1µA
0.02%RD + 1.1 μA
Resistance
1 to 400.00 Ω
1 to 4000.0 Ω
10mΩ
100mΩ
0.02%RD + 0.02 Ω
0.03%RD + 0.4 Ω
Frequency
0 to 50000.0 Hz
0.1Hz
0.005%RD + 1 Hz
Pulse
0 to 999999
1
N/A
DC24V
N/A
N/A
0.5V

• General Specifications

Environmental Specifications
Operating Temperature
-10°C to 50°C
Storage Temperature
-20°C to 60°C
Humidity
<90%, non-condensing
Safety Specifications
European Compliance
CE Mark
Mechanical Specifications
Display
3.5 inch TFT color screen
Electrical Connection
φ4mm sockets and flat mini-jack thermocouple socket
RS232 Interface
standard RS232-DB9 socket
Size
3.9” x 7.6” x 2.0” (100mm x 192mm x 52mm)
Weight
1.6 lb (0.7 kg)
Power Supply
Polymer Li-ion rechargeable battery, or 10V DC adaptor
Battery
Rechargeable Li-ion battery (included)
Battery Life
15 hours uninterrupted use
Battery life will be reduced when 24V is applied
Battery Charge
110V/220V external power adapter (included)
Pressure Specification
   The 160 series Intelligent Digital Pressure Modules are available for gauge, vacuum and absolute pressure from -15 psi to 10,000 psi (-1 bar to 700 bar). Accuracy from 0.025% FS includes operation over 14°F to 122°F (-10°C to 50°C), one year stability and calibration uncertainty. For detail specification refer to pressure modules datasheet.
 

• Temperature Specifications

Thermocouple Measurement and Source accuracy
Measure
and Simulate
Standard
Temperature Range (°C)
Accuracy (°C)
Measure
Source
S
IEC 584
-50 to 1768
-50 to400
1.0
1.1
400 to 1000
0.6
0.6
1000 to1768
0.7
0.8
R
IEC 584
-50 to 1768
-50 to 200
1.4
1.4
200 to 500
0.6
0.6
500 to 1768
0.6
0.7
B
IEC 584
0 to 1820
50 to 450
3.8
3.8
450 to 800
0.9
0.9
800 to 1820
0.6
0.7
K
IEC 584
-270 to 1372
-250 to -200
1.0
1.1
-200 to -100
0.4
0.5
-100 to 600
0.3
0.3
600 to 1372
0.4
0.5
N
 IEC 584
-270 to 1300
-250 to -200
1.5
1.6
-200 to -100
0.5
0.6
-100 to 1300
0.4
0.5
E
IEC 584
-270 to 1000
-250 to -200
0.6
0.7
-200 to -100
0.3
0.3
-100 to 0
0.2
0.2
0 to 700
0.2
0.3
700 to 1000
0.2
0.4
J
IEC 584
-270 to 1200
-210 to -100
0.3
0.3
-100 to 1200
0.3
0.4
T
IEC 584
-270 to 400
-250 to -200
0.8
0.9
-200 to 0
0.4
0.4
0 to 400
0.2
0.2
C
ASTM E988
0 to 2315
0 to 1000
0.5
0.5
1000 to 1800
0.7
0.9
1800 to 2315
1.0
1.4
D
ASTM E988
0 to 2320
0 to 100
0.5
0.5
100 to 1100
0.4
0.5
1100 to 2000
0.6
0.9
2000 to 2320
0.9
1.3
G
ASTM E1751
0 to 2315
0 to 200
2.4
2.4
200 to 400
0.5
0.5
400 to 1400
0.4
0.5
1400 to 2315
0.7
1.0
L
DIN43710
-200 to 900
-200 to -100
0.2
0.3
-100 to 400
0.2
0.2
400 to 900
0.2
0.3
U
DIN43710
-200 to 600
-200 to 0
0.4
0.4
0 to 600
0.2
0.3
RTD Measurement and Source accuracy
Measure
and Simulate
Standard
Temperature Range (°C)
Accuracy (°C)
Measure (2W/3W)
Measure (4W)
Source
Pt10(385)
IEC 751
-200 to 850
-100 to 200
0.65
0.60
0.65
200 to 600
0.82
0.72
0.82
600 to 850
0.96
0.82
0.96
PT100
(385)
IEC 751
-200 to 850
-100 to 200
0.15
0.1
0.15
200 to 600
0.26
0.16
0.26
600 to 850
0.34
0.20
0.34
Pt100
(3916)
IEC 751
-200 to 850
-100 to 200
0.15
0.1
0.15
200 to 600
0.26
0.16
0.26
600 to 850
0.33
0.20
0.33
Pt200(385)
IEC 751
-200 to 850
-100 to 200
0.37
0.32
0.69
200 to 600
0.51
0.41
0.92
600 to 850
0.61
0.48
1.08
Pt500(385)
IEC 751
-200 to 850
-100 to 200
0.20
0.16
0.36
200 to 600
0.32
0.22
0.54
600 to 850
0.40
0.27
0.67
Pt1000
(385)
IEC 751
-200 to 850
-100 to 200
0.1
0.05
0.26
200 to 600
0.2
0.10
0.42
600 to 850
0.27
0.14
0.54
Cu100(428)
GOST6651-94
-50 to 150
-50 to150
0.12
0.09
0.12
Cu50(428)
GOST6651-94
-50 to 150
-50 to 150
0.17
0.13
0.17
Cu10(427)
Minco Application Aid #18
-100 to 260
-100 to 260
0.61
0.56
0.61
Ni120(672)
Edison curve #7
-100 to 260
-100 to 260
0.07
0.05
0.07
Ni100(618)
DIN 43760
-100 to 260
-100 to 260
0.08
0.06
0.08
• ORDERING INFORMATION:
      Model Number:  ADT223A

Accessories (included)

110V/220V external power adapter
1 pc
Chargeable Li-ion battery
1 pc
 Test leads
2 sets (4 pcs)
short circuit cable
1 sets (2 pcs)
Manual
1 pc
NIST traceable calibration certificate
1 pc
 

Optional Accessories

Model number
Description
ADT 160
Intelligent Digital Pressure Modules
9060
Pressure module connection cable
9050
USB to RS232 (DB-9 Male) Adapter
9080
Cold Junction compensation kits (including TC plug, compensation cable, S,R,B,K,J,T,E,N)
9712
Spare chargeable Li-ion battery for multifunction calibrator
9816
110V/220V external power adapter for multifunction calibrator
9906
Carrying case for multifunction calibrator
9510
Additel/Cal Task management software for multifunction calibrator

 

• Electrical Specifications

Measurement Accuracy
Range
Resolution
Accuracy 
Voltage DC
±75.0000 mV
±30.0000 V
0.11µV
0.1mV
0.01%RD+3.75 μV
0.01%RD+1.5 mV
Current DC
±30.0000 mA
0.1µA
0.01%rdg + 1.5 μA
Resistance
Two-wire
0 to 400.000 Ω
1mΩ
0.02%RD + 0.02 Ω
Three-wire
0 to 400.000 Ω
1mΩ
0.02%RD + 0.02 Ω
Four-wire
0 to 400.000 Ω
1mΩ
0.01%RD + 0.02 Ω
Two-wire
0 to 400.000 Ω
10mΩ
0.02%RD + 0.2 Ω
Three-wire
0 to 400.000 Ω
10mΩ
0.02%RD + 0.2 Ω
Four-wire
0 to 400.000 Ω
10mΩ
0.01%RD + 0.2 Ω
Frequency
1 to 50000.0 Hz
0.1Hz
0.005%RD + 1 Hz
Pulse
0 to 999999
1
N/A
Limit Switch
If the switch has detective voltage, its range is +3V to +24V
Source Accuracy
Voltage DC
-10.000 to 75.000mV 0 to 12.0000 V
1μV
0.1mV
0.02%RD + 4.25 μV
0.02%RD + 0.6 mV
Current DC
0 to 22.000 mA
1µA
0.02%RD + 1.1 μA
Resistance
1 to 400.00 Ω
1 to 4000.0 Ω
10mΩ
100mΩ
0.02%RD + 0.02 Ω
0.03%RD + 0.4 Ω
Frequency
0 to 50000.0 Hz
0.1Hz
0.005%RD + 1 Hz
Pulse
0 to 999999
1
N/A
DC24V
N/A
N/A
0.5V

• General Specifications

Environmental Specifications
Operating Temperature
-10°C to 50°C
Storage Temperature
-20°C to 60°C
Humidity
<90%, non-condensing
Safety Specifications
European Compliance
CE Mark
Mechanical Specifications
Display
3.5 inch TFT color screen
Electrical Connection
φ4mm sockets and flat mini-jack thermocouple socket
RS232 Interface
standard RS232-DB9 socket
Size
3.9” x 7.6” x 2.0” (100mm x 192mm x 52mm)
Weight
1.6 lb (0.7 kg)
Power Supply
Polymer Li-ion rechargeable battery, or 10V DC adaptor
Battery
Rechargeable Li-ion battery (included)
Battery Life
15 hours uninterrupted use
Battery life will be reduced when 24V is applied
Battery Charge
110V/220V external power adapter (included)
Pressure Specification
   The 160 series Intelligent Digital Pressure Modules are available for gauge, vacuum and absolute pressure from -15 psi to 10,000 psi (-1 bar to 700 bar). Accuracy from 0.025% FS includes operation over 14°F to 122°F (-10°C to 50°C), one year stability and calibration uncertainty. For detail specification refer to pressure modules datasheet.
 

• Temperature Specifications

Thermocouple Measurement and Source accuracy
Measure
and Simulate
Standard
Temperature Range (°C)
Accuracy (°C)
Measure
Source
S
IEC 584
-50 to 1768
-50 to400
1.0
1.1
400 to 1000
0.6
0.6
1000 to1768
0.7
0.8
R
IEC 584
-50 to 1768
-50 to 200
1.4
1.4
200 to 500
0.6
0.6
500 to 1768
0.6
0.7
B
IEC 584
0 to 1820
50 to 450
3.8
3.8
450 to 800
0.9
0.9
800 to 1820
0.6
0.7
K
IEC 584
-270 to 1372
-250 to -200
1.0
1.1
-200 to -100
0.4
0.5
-100 to 600
0.3
0.3
600 to 1372
0.4
0.5
N
 IEC 584
-270 to 1300
-250 to -200
1.5
1.6
-200 to -100
0.5
0.6
-100 to 1300
0.4
0.5
E
IEC 584
-270 to 1000
-250 to -200
0.6
0.7
-200 to -100
0.3
0.3
-100 to 0
0.2
0.2
0 to 700
0.2
0.3
700 to 1000
0.2
0.4
J
IEC 584
-270 to 1200
-210 to -100
0.3
0.3
-100 to 1200
0.3
0.4
T
IEC 584
-270 to 400
-250 to -200
0.8
0.9
-200 to 0
0.4
0.4
0 to 400
0.2
0.2
C
ASTM E988
0 to 2315
0 to 1000
0.5
0.5
1000 to 1800
0.7
0.9
1800 to 2315
1.0
1.4
D
ASTM E988
0 to 2320
0 to 100
0.5
0.5
100 to 1100
0.4
0.5
1100 to 2000
0.6
0.9
2000 to 2320
0.9
1.3
G
ASTM E1751
0 to 2315
0 to 200
2.4
2.4
200 to 400
0.5
0.5
400 to 1400
0.4
0.5
1400 to 2315
0.7
1.0
L
DIN43710
-200 to 900
-200 to -100
0.2
0.3
-100 to 400
0.2
0.2
400 to 900
0.2
0.3
U
DIN43710
-200 to 600
-200 to 0
0.4
0.4
0 to 600
0.2
0.3
RTD Measurement and Source accuracy
Measure
and Simulate
Standard
Temperature Range (°C)
Accuracy (°C)
Measure (2W/3W)
Measure (4W)
Source
Pt10(385)
IEC 751
-200 to 850
-100 to 200
0.65
0.60
0.65
200 to 600
0.82
0.72
0.82
600 to 850
0.96
0.82
0.96
PT100
(385)
IEC 751
-200 to 850
-100 to 200
0.15
0.1
0.15
200 to 600
0.26
0.16
0.26
600 to 850
0.34
0.20
0.34
Pt100
(3916)
IEC 751
-200 to 850
-100 to 200
0.15
0.1
0.15
200 to 600
0.26
0.16
0.26
600 to 850
0.33
0.20
0.33
Pt200(385)
IEC 751
-200 to 850
-100 to 200
0.37
0.32
0.69
200 to 600
0.51
0.41
0.92
600 to 850
0.61
0.48
1.08
Pt500(385)
IEC 751
-200 to 850
-100 to 200
0.20
0.16
0.36
200 to 600
0.32
0.22
0.54
600 to 850
0.40
0.27
0.67
Pt1000
(385)
IEC 751
-200 to 850
-100 to 200
0.1
0.05
0.26
200 to 600
0.2
0.10
0.42
600 to 850
0.27
0.14
0.54
Cu100(428)
GOST6651-94
-50 to 150
-50 to150
0.12
0.09
0.12
Cu50(428)
GOST6651-94
-50 to 150
-50 to 150
0.17
0.13
0.17
Cu10(427)
Minco Application Aid #18
-100 to 260
-100 to 260
0.61
0.56
0.61
Ni120(672)
Edison curve #7
-100 to 260
-100 to 260
0.07
0.05
0.07
Ni100(618)
DIN 43760
-100 to 260
-100 to 260
0.08
0.06
0.08
• ORDERING INFORMATION:
      Model Number:  ADT223A

Accessories (included)

110V/220V external power adapter
1 pc
Chargeable Li-ion battery
1 pc
 Test leads
2 sets (4 pcs)
short circuit cable
1 sets (2 pcs)
Manual
1 pc
NIST traceable calibration certificate
1 pc
 

Optional Accessories

Model number
Description
ADT 160
Intelligent Digital Pressure Modules
9060
Pressure module connection cable
9050
USB to RS232 (DB-9 Male) Adapter
9080
Cold Junction compensation kits (including TC plug, compensation cable, S,R,B,K,J,T,E,N)
9712
Spare chargeable Li-ion battery for multifunction calibrator
9816
110V/220V external power adapter for multifunction calibrator
9906
Carrying case for multifunction calibrator
9510
Additel/Cal Task management software for multifunction calibrator

 

  • Measure and source 13 T/Cs, 11 RTDs, mV, mA, Ohms, frequency and pulses, with accuracy to 0.01% of reading
  • Pressure via
  • external pressure modules
  • up to 70,000kPa
  • 24V loop power
  • HART communication capability
  • Task management capability, upload, download and document
  • Easy to use smartphone type menu and interface
  • Rugged design for use in harsh environments