Series 2-5000 Minihelic® II Differential Pressure Gage

Minihelic® II Differential Pressure Gage Combining High Accuracy, Compactness, Dependability, and Low Cost
Combining High Accuracy, Compactness, Dependability, and Low Cost

Combining clean design, small size and low cost with enough accuracy for all but the most demanding applications our SERIES 2-5000 Minihelic® II gage offers the latest in design features for a dial type differential pressure gage. It is our most compact gage but is easy to read and can safely operate at total pressures up to 30 psig.

Product Applications

  • Room Positive Pressure Sensing
  • Local Indication on Filter Status
  • Face Velocity on Fume Hood
  • Duct Pressures
  • Cabinet Air-Purging
  • Medical Respiratory Equipment
  • Air Samplers
  • Electronic Air Cooling Systems
  • Laminar Flow Hoods
More Information
Model Chart
EXAMPLE 2-5000-0 AT Series 2-5000-0-AT Minihelic® II Differential Pressure Gage, range 0-.5" w.c., with aluminum tag.
RANGE 2-5000-0
2-5001
2-5002
2-5003
2-5005
2-5010
2-5020
2-5040
2-5060
2-5100
2-5200
2-5205
2-5210
2-5215
2-5000-25MM
2-5000-50MM
2-5000-100MM
2-5000-125PA
2-5000-250PA
2-5000-500PA
2-5000-750PA
2-5000-1KPA
2-5000-3KPA
2-5000-5KPA
  0-.5" w.c.
0-1" w.c.
0-2" w.c.
0-3" w.c.
0-5" w.c.
0-10" w.c.
0-20" w.c.
0-40" w.c.
0-60" w.c.
0-100" w.c.
0-200" w.c.
0-5 psid
0-10 psid
0-15 psid
0-25 mm w.c.
0-50 mm w.c.
0-100 mm w.c.
0-125 Pa
0-250 Pa
0-500 Pa
0-750 Pa
0-1 kPa
0-3 kPa
0-5 kPa
OPTIONS   AT
BB
NIST
NPT
Aluminum Tag
Bottom Barb Connection
NIST Certificate of Calibration
NPT Connection
Service:Air and compatible gases.
Wetted Materials:Consult factory.
Housing:Glass filled nylon; acrylic lens.
Accuracy:±5% of FS at 70°F (21.1°C).
Pressure Limits:30 psig (2.067 bar) continuous to either pressure connection.
Temperature Limits:20 to 120°F (-6.67 to 48.9°C).
Mounting Orientation:Diaphragm in vertical position. Consult factory for other position orientations.
Process Connections:Barbed, for 3/16" ID tubing (standard); 1/8" male NPT (optional).
Weight:6 oz (170.1g).
Agency Approvals:Meets the technical requirements of EU Directive 2011/65/EU (RoHS II).
CAUTION: FOR USE ONLY WITH AIR OR COMPATIBLE GASES.
  • Removable lens and rear-housing provides easy, cost-effective servicing
  • Accuracy and value provides an excellent solution for OEM and user applications
  • Durable housing materials make it well-suited for rough environments and total high pressure