Differential pressure gauge with capsule element (Price & availability on application)

Differential pressure gauge with capsule element (Price & availability on application)
Design
For very low differential pressures, DT - GM 87 10 226

Nominal size in mm
Model 716.11: NS 63, 100, 160
Model 736.11: NS 100, 160

Accuracy class
1.6

Scale ranges
Model 716.11:
NS 63: 0 ... 16 to 0 ... 400 mbar
NS 100: 0 ... 10 to 0 ... 250 mbar
NS 160: 0 ... 6 to 0 ... 250 mbar

Model 736.11:
NS 100: 0 ... 16 to 0 ... 250 mbar
NS 160: 0 ... 1.6 to 0 ... 250 mbar

or all other equivalent vacuum or combined pressure and vacuum ranges

Pressure limitation
Steady: Full scale value
Fluctuating: 0.9 x full scale value

Overpressure safety
Full scale value

Max. working pressure (static pressure)
NS 63: 400 mbar
NS 100, 160: 250 mbar

Permissible temperature
Ambient: -20 ... +60 °C
Medium: +70 °C maximum

Temperature effect
When the temperature of the measuring system deviates from the reference temperature (+20 °C):
max. ±0.5 %/10 K of full scale value

Ingress protection
IP 66 per EN 60529 / lEC 529
Product Description

Details

Design For very low differential pressures, DT - GM 87 10 226 Nominal size in mm Model 716.11: NS 63, 100, 160 Model 736.11: NS 100, 160 Accuracy class 1.6 Scale ranges Model 716.11: NS 63: 0 ... 16 to 0 ... 400 mbar NS 100: 0 ... 10 to 0 ... 250 mbar NS 160: 0 ... 6 to 0 ... 250 mbar Model 736.11: NS 100: 0 ... 16 to 0 ... 250 mbar NS 160: 0 ... 1.6 to 0 ... 250 mbar or all other equivalent vacuum or combined pressure and vacuum ranges Pressure limitation Steady: Full scale value Fluctuating: 0.9 x full scale value Overpressure safety Full scale value Max. working pressure (static pressure) NS 63: 400 mbar NS 100, 160: 250 mbar Permissible temperature Ambient: -20 ... +60 °C Medium: +70 °C maximum Temperature effect When the temperature of the measuring system deviates from the reference temperature (+20 °C): max. ±0.5 %/10 K of full scale value Ingress protection IP 66 per EN 60529 / lEC 529
Additional Information

Additional Information

Applications Differential pressure measurement at measuring points with very low differential pressures, for gaseous, dry, clean, oil and grease free media | Model 736.11 also for aggressive media and environment | Filter monitoring in ventilation and heating systems | Filter monitoring in overpressure and clean rooms | Differential pressure controlled monitoring of ventilator and blast pressures
Description Design For very low differential pressures, DT - GM 87 10 226 Nominal size in mm Model 716.11: NS 63, 100, 160 Model 736.11: NS 100, 160 Accuracy class 1.6 Scale ranges Model 716.11: NS 63: 0 ... 16 to 0 ... 400 mbar NS 100: 0 ... 10 to 0 ... 250 mbar NS 160: 0 ... 6 to 0 ... 250 mbar Model 736.11: NS 100: 0 ... 16 to 0 ... 250 mbar NS 160: 0 ... 1.6 to 0 ... 250 mbar or all other equivalent vacuum or combined pressure and vacuum ranges Pressure limitation Steady: Full scale value Fluctuating: 0.9 x full scale value Overpressure safety Full scale value Max. working pressure (static pressure) NS 63: 400 mbar NS 100, 160: 250 mbar Permissible temperature Ambient: -20 ... +60 °C Medium: +70 °C maximum Temperature effect When the temperature of the measuring system deviates from the reference temperature (+20 °C): max. ±0.5 %/10 K of full scale value Ingress protection IP 66 per EN 60529 / lEC 529
Model No 716.11, measuring system copper alloy |736.11, measuring system stainless steel
Special Features Differential pressure measuring ranges from 0 … 4 mbar | As a standard zero point adjustable in front | Ingress protection IP 66 | Case and pressur element made of stainless steel
Download Brochures
Download Attachment
pdfPDF File 1(265.39 KB )Download
pdfPDF File 2(314.11 KB )Download